Bioprosthetic Valve Fracture During Valve-in-valve TAVR: Bench to Bedside

Register or Login to View PDF Permissions
Permissions× For commercial reprint enquiries please contact Springer Healthcare: ReprintsWarehouse@springernature.com.

For permissions and non-commercial reprint enquiries, please visit Copyright.com to start a request.

For author reprints, please email rob.barclay@radcliffe-group.com.
Average (ratings)
No ratings
Your rating

Abstract

Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) has been established as a safe and effective means of treating failed surgical bioprosthetic valves (BPVs) in patients at high risk for complications related to reoperation. Patients who undergo VIV TAVR are at risk of patient–prosthesis mismatch, as the transcatheter heart valve (THV) is implanted within the ring of the existing BPV, limiting full expansion and reducing the maximum achievable effective orifice area of the THV. Importantly, patient–prosthesis mismatch and high residual transvalvular gradients are associated with reduced survival following VIV TAVR. Bioprosthetic valve fracture (BVF) is as a novel technique to address this problem. During BPV, a non-compliant valvuloplasty balloon is positioned within the BPV frame, and a highpressure balloon inflation is performed to fracture the surgical sewing ring of the BPV. This allows for further expansion of the BPV as well as the implanted THV, thus increasing the maximum effective orifice area that can be achieved after VIV TAVR. This review focuses on the current evidence base for BVF to facilitate VIV TAVR, including initial bench testing, procedural technique, clinical experience and future directions.

Disclosure:Dr. Chhatriwalla receives research and clinical trial support from, acts as a proctor of, and is on the speakers bureau of Medtronic, Edwards Lifesciences and St. Jude Medical. Dr. Cohen receives research and clinical trial support from Medtronic, St. Jude Medical, and Edwards Lifesciences. Dr. Allen receives research support from Abbott Medical and Edwards Lifesciences, and is on the speakers bureau of Edwards Lifesciences and Medtronic. Dr. Saxon has no disclosures.

Received:

Accepted:

Correspondence Details:Adnan Chhatriwalla, MD, Saint Luke’s Mid America Heart Institute, 4330 Wornall Road, Suite 2000, Kansas City, MO 64111, USA. E: achhatriwalla@saint-lukes.org

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

Surgical aortic valve replacement (SAVR) has long been the standard of care for aortic valve replacement. More recently, transcatheter aortic valve replacement (TAVR) has been established as a safe, effective and less invasive method of valve replacement in patients with severe aortic stenosis who are at intermediate or high risk for complications related to SAVR.1–4 For patients who present with surgical bioprosthetic valve (BPV) degeneration, reoperation may be associated with increased risks.5,6 Valve-in-valve (VIV) TAVR has emerged as a safe and effective therapy for such patients.7,8 The US Food and Drug Administration has approved VIV TAVR with both self-expanding and balloon-expandable prostheses for patients with failed BPVs who are at high risk for complications related to reoperation.

Although VIV TAVR is a promising alternative to repeat SAVR, patient–prosthesis mismatch (PPM) is a concern. Broadly, PPM is defined as any situation in which the effective orifice area (EOA) of a prosthetic valve is smaller than the orifice of the patients’ native aortic valve; severe aortic PPM is defined as an indexed EOA ≤0.65 cm2/m2.9 Patients who undergo VIV TAVR are particularly at risk for PPM because the TAVR prosthesis is implanted within the frame of the previous BPV, thereby reducing the maximum EOA that can be achieved with the new valve. Importantly, severe PPM and small labelled BPV size (≤21 mm) have been associated with higher mortality following VIV TAVR.10–12 In the VIVID Registry, which is the largest series of VIV TAVR published to date, the incidence of severe PPM following VIV TAVR was 31.8 %.7 Furthermore, patients in the same series with a small labelled surgical valve size (≤21 mm) had a reduced survival at 1 year (74.8 %) compared with patients with an intermediate (21–25 mm) sized BPV (81.8 %) or a large (≥25 mm) BPV (93.3 %), suggesting that PPM has a negative impact on survival following VIV TAVR.

Several strategies have been developed to avoid PPM following VIV TAVR. Use of a transcatheter heart valve (THV) with supra-annular leaflet positioning (e.g. CoreValve Evolut, Medtronic, Minneapolis, MN, USA) may result in a larger EOA due to the supra-annular position of the prosthetic leaflets, and a higher THV implant depth may improve inflow dynamics and augment the effective EOA.13–17 Recently, several publications have reported on the concept of fracturing the surgical BPV ring with a high-pressure balloon inflation in order to dilate the BPV and permit further expansion of the THV,18–20 improving haemodynamic results in such patients.21–24 This review summarises our current knowledge of BPV fracture (BVF) to facilitate VIV TAVR, including information obtained from bench testing, procedural techniques, early clinical experience and future directions.

Bench Testing

To date, there have been two published series regarding the tolerance of commercially available BPVs to high-pressure balloon inflation using non-compliant valvuloplasty balloons. Results of BVF in Trifecta (St Jude, Minneapolis, MN, USA), Mitroflow (Sorin, Milan, Italy), Magna Ease (Edwards Lifesciences, Irvine, CA, USA), Mosaic (Medtronic), Magna (Edwards Lifesciences), Hancock II (Medtronic) and Biocor Epic (St Jude) valves using Atlas Gold (Bard, Tempe, AZ, USA) and True Balloons (Bard) have been reported.21,24

The procedural technique for BVF bench testing is displayed in Figure 1. Non-compliant balloons were positioned within small BPVs (i.e. labelled valve size 19 or 21 mm). Valvuloplasty balloons were sized 1 mm larger than the labelled surgical valve size. A high-pressure stopcock was used to separately attach a syringe and an indeflator to the balloon. With the stopcock open to the syringe, an initial hand inflation was performed to rapidly inflate the balloon, then the stopcock was opened to the indeflator, and the pressure was gradually increased in the balloon system until the BPV ring fractured. Fracture of the BPV ring was typically associated with a sudden decrease in inflation pressure, visible release of the balloon waist, and/or an audible “click”. BVF was considered unsuccessful if the balloon ruptured without evidence of valve fracture. The fracture threshold was reported as the lowest inflation pressure necessary to cause BPV ring fracture. Following BVF, the fractured valves were inspected for protruding elements or other potentially harmful results related to BPV disruption.21,24

The findings of systematic bench testing are summarised in Table 1. BPVs that could be consistently fractured in both series were Magna, Magna Ease, Mitroflow, Mosaic and Biocor Epic. Upon dissection of the sewing cuff, the fractured elements were directly visualised to confirm complete separation of the ring element (Figure 2).21,24 Hancock II valves could not be fractured in either series. BVF of 19 mm Trifecta valves was also unsuccessful in both series.21 A partial fracture of the 21 mm Trifecta valve (separation of the lower of two parallel rings of the frame) was noted by both groups. However, this partial fracture only occurred at very high inflation pressure (26 atm) or after the use of serial balloon inflations with increasing balloon diameter. Neither authors recommended BVF in Trifecta valves.

The minimum inflation pressures necessary to fracture similar valves were slightly different in the two series (Table 1). In general terms, BPVs with an alloyed metal ribbon ring (Magna and Magna Ease) demonstrated a higher fracture threshold (18–24 atm) than BPVs with a polymer ring (Biocor Epic, Mosaic, Mitroflow; 8–12 atm). Different techniques were used to measure inflation pressure at the moment of fracture in the two series, thus the small differences in observed fracture thresholds were not unexpected. In practical terms, precise knowledge of the fracture threshold may not be necessary. Rather, when a clinical case of BVF is planned, the most critical information is the knowledge that a particular type of valve can be fractured, and the approximate inflation pressure that will result in fracture. Therefore, the largely concordant findings between the two series serves as an excellent guide for operators when planning a BVF procedure.

Indications

At present, the indications to perform BVF to facilitate VIV TAVR are not fully defined. The majority of patients, in particular those with large surgical valves, are likely to achieve an adequate haemodynamic result with VIV TAVR, and patients without PPM following VIV TAVR have an excellent survival to 1 year.7,8 Therefore, patients who stand to benefit the most from BVF are those who are predisposed to PPM and high residual transvalvular gradients following VIV TAVR, including those with small BPVs (labelled valve size ≤21 mm) and/or stenosis as the mechanism of BPV failure.7,8 Whether patients with large BPVs (>21 mm labelled valve size) or intermediate transvalvular gradients (10–20 mmHg) after VIV TAVR stand to benefit from BVF is not known.

Clinical Application

Translating the ex vivo BVF technique to a clinical setting is intuitive (Figure 2). During a case of VIV TAVR, a non-compliant valvuloplasty balloon, such as those used in bench testing, is positioned across the BPV ring (Figure 3a) over a stiff wire. At that point, the procedural technique is the same as depicted in Figure 1b,c. During rapid ventricular pacing, an initial hand inflation is used to fill the balloon with dilute contrast, then a coronary indeflator is used to increase the inflation pressure to the threshold for valve fracture. BPV fracture is accompanied by the same auditory, visual and haptic feedback as is observed during bench testing: a sudden drop in inflation pressure, a visible release of the balloon waist (Figure 3b,c) and/or an audible “click”. The valvuloplasty balloon is then deflated and removed. Figure 3d depicts the final result from the same clinical case, with a dramatic increase in expansion of both the TAVR prosthesis and BPV ring. Of note, given the prolonged nature of the pacing run that is often required, we prefer to perform all BVF procedures under general anesthesia to minimise any temporary neurologic sequelae.

Figure 1: Technique of High-pressure Balloon Inflation to Perform Bioprosthetic Valve Fracture

Article image

An example of the haemodynamic result of a clinical case of BVF with VIV TAVR is depicted in Figure 3. A 76-year-old man with a history of combined coronary artery bypass grafting (CABG) and SAVR, with a 23 mm Mosaic BPV 10 years prior, presented with symptoms of class III diastolic heart failure and severe BPV stenosis, with a mean transvalvular gradient of 49 mmHg, a peak Doppler velocity of 4.7 m/s and dimensionless index of 0.25. The Society of Thoracic Surgeons (STS) predicted risk of mortality was 8.2 %, and after a Heart Team evaluation, VIV TAVR was recommended. Baseline invasive haemodynamics (Figure 3a) demonstrated a mean gradient of 36 mmHg with a calculated valve EOA of 0.7 cm2. After deployment of a 26 mm CoreValve Evolut R self-expanding TAVR prosthesis, the mean gradient was reduced to 25 mmHg, with a corresponding EOA of 1.2 cm2 (Figure 3b). BVF was then performed with a 24 mm True Balloon, and the bioprosthetic ring fractured at 10 atm. Final haemodynamics demonstrated a mean gradient of 3 mmHg and an EOA of 1.7 cm2 (Figure 3c). At 1 month follow up, the patient was doing well, with New York Heart Association class 1 functional status. An echocardiogram at that time demonstrated a mean transvalvular gradient of 8 mmHg, with a peak Doppler velocity of 2.0 m/s and a dimensionless index of 0.68.

Clinical Experience

The procedural and haemodynamic results of patients who have been treated with VIV TAVR and BVF in two published case series are displayed in Table 2.22,23 A total of 30 patients with a mean age of 79.0 years were treated with VIV TAVR for failed BPVs. The majority of cases were performed to treat BPV stenosis, with a mean time from SAVR implant to VIV TAVR of 10.4 years. Fifteen patients were treated with TAVR prior to BVF, and 15 patients were treated with BVF followed by TAVR implant. There were no reports of perioperative death, coronary artery obstruction, annular rupture, aortic root injury, paravalvular leak or pericardial effusion. Two patients suffered small periprocedural strokes that were confirmed with MRI, and both patients later recovered full neurological function.22,23 Ten cases were performed with backup haemodynamic support with extracorporeal membrane oxygenation (ECMO), and 100 % long-term survival was noted in one case series (n=10), with a mean follow up of 438 days.23

Table 1: Combined Results of Bioprosthetic Valve Fracture Bench Testing

Article image

The procedural results reported in these two case series highlight the haemodynamic benefit of BVF (Table 2). For the combined cohort, the mean gradient was reduced from 41 mmHg pre-procedure to 11 mmHg after BVF and VIV TAVR, which corresponds to an improvement in EOA from 0.75 to 1.7 cm2. In one series, most of the patients (15/20) were treated with BVF after VIV TAVR was performed. For this subset of patients, the mean pre-procedural gradient was reduced from 41.9 to 20.5 mmHg after VIV TAVR, and the mean gradient was further reduced from 20.5 to 6.7 mmHg following BVF. This corresponds to mean EOAs of 0.6, 1.0 and 1.7 cm2, respectively. The benefit of BVF to improve the procedural results of VIV TAVR is evident: with VIV TAVR alone, these patients would have been left with a suboptimal EOA of 1.0 cm2 and a final mean gradient of 20.5 mmHg.

Complications

Although the haemodynamic benefit of BVF is clear, the incremental risk posed by BVF in addition to VIV TAVR is not fully known. In the two clinical series published to date (total n=30), complications were few. Two patients suffered a small periprocedural stroke, with complete resolution of neurological deficits. One patient suffered complete atrioventricular block requiring a permanent pacemaker.22,23 However, stroke and heart block are potential complications of TAVR even in the absence of BVF. Despite the relatively low complication rate in the published series and unpublished clinical experience, there are some theoretical risks specific to the BVF procedure that must be considered. Although there were no incidents of aortic root disruption, paravalvular leak, aortic insufficiency or coronary occlusion in the published case series, it is important to acknowledge that the clinical experience with BVF is still early. There is still much to learn about the specific clinical and anatomic features that predispose patients to complications.

Coronary Artery Obstruction

As is true of VIV TAVR, one of the major concerns with BVF is the potential for coronary artery obstruction. A recent registry of VIV TAVR reported an incidence of coronary artery obstruction of 3.5 % with VIV TAVR,25 which appears to be decreasing as experience with VIV TAVR grows.26 There are several risk factors for coronary obstruction during VIV TAVR: narrow coronary sinuses, low coronary artery ostia, bulky bioprosthetic leaflets, reimplanted coronary arteries and type of BPV, i.e. those with leaflets mounted external to the stent frame (Mitroflow, Trifecta).26 Whereas the typical concern with native valve TAVR is coronary ostial height in relation to the native aortic annulus, during VIV TAVR the most important factor is the anticipated distance between the coronary ostia and the final position of the BPV leaflets.26 This relationship can be assessed during pre-procedural coronary angiography as well as by computed tomography, wherein a “virtual THV” can be superimposed on the CT images to determine the relationship between the BPV leaflets and the coronary arteries.26 A virtual THV to coronary distance of less than 3 mm is considered to place a patient at high risk for coronary occlusion.26

With BVF, the architecture of the BPV is altered such that the final position of the BPV leaflets is less certain. In bench testing, measurement of the BPV following BVF demonstrated a maximum gain in BPV diameter of 3–4 mm. Further expansion of the BPV is restricted by the Dacron sewing cuff, which remains intact after valve fracture.21 The additional space in the coronary sinuses necessary to accommodate BVF is not fully understood. Extrapolating from the recommended safety margins of VIV TAVR, it is reasonable to estimate that a BPV to coronary distance of less than 5 mm could be considered to place a patient at high risk for coronary occlusion when BVF is performed. To date, there are no published cases of coronary occlusion attributable to BVF. If there is pre-procedural concern for coronary occlusion, standard precautions are recommended, including wire protection of the coronary artery as deemed appropriate by the operators.

THV selection

Careful selection of the THV is an important part of the evaluation prior to BVF. Both self-expanding and balloon-expandable THVs are currently approved for use in VIV TAVR in the US. There are some data to suggest that self-expanding THVs result in superior procedural haemodynamics and increased EOA when used for VIV TAVR, compared with balloon-expandable valves,14,17 largely due to the supra-annular position of the prosthetic leaflets on the self-expanding frame. However, BVF can be successfully performed in the setting of VIV with both self-expanding and balloon-expandable THVs. There is some concern that the high-pressure balloon inflation used to perform BVF may cause structural damage to the self-expanding valve frame or leaflets, resulting in severe acute valvular regurgitation. This can largely be avoided by using a balloon smaller than the constrained segment of the self-expanding THV, and by positioning the BVF balloon such that the balloon shoulder is lower (i.e. more ventricular) than where the leaflets are anchored to the frame.21 However, care must be employed to ensure the valvuloplasty balloon and delivery wire are well-positioned in the ventricle as well.

In terms of TAVR prosthesis selection, BVF adds an extra element of pre-procedural consideration. Appropriate THV selection for VIV TAVR is guided by the true inner diameter of the BPV, rather than the labelled surgical valve size, as there can be considerable difference in these measurements between different valve models.27 BVF results in structural expansion of the BPV, changing the true inner diameter of the BPV considerably. Based on measurements made during bench testing, it appears that the maximum gain in diameter that can be achieved with BVF is between 3 and 4 mm.21 Thus, there are some situations in which a larger TAVR prosthesis may be appropriate for a BVF procedure than would be selected for a stand-alone VIV TAVR. For example, in a patient with a 21 mm Mitroflow, which has an inner diameter of 17 mm, BVF might result in expansion of the inner diameter to 20–21 mm. In this situation, it is not known whether a partially constrained 23 mm transcatheter valve would result in better haemodynamics than a fully expanded 20 mm transcatheter valve. In vitro testing has suggested that a larger prosthesis, even if expanded to a less than nominal diameter, may result in a more favourable transvalvular gradient.17 However, this concept has not been rigorously tested in clinical practice, and the interaction between TAVR prosthesis expansion and optimal haemodynamics after BVF is not fully understood.

Figure 2: Fractured 21 mm Mitroflow Bioprosthetic Valve

Article image

Selection of Valvuloplasty Balloon

Both Atlas Gold and True Dilatation balloons are consistently able to fracture small surgical valves, both in bench testing (Table 1) and clinical experience.21–24 In bench testing and the majority of clinical cases, balloons sized 1 mm larger than the labelled valve size were utilised. However, during some clinical cases, smaller balloons (i.e. sized 1 mm larger than the true inner diameter of the BPV) were used successfully. It stands to reason that balloons need only be sized larger than the internal diameter of the BPV to fracture the valve, especially if a THV is already implanted in the BPV. Whether smaller balloons consistently fracture BPVs remains to be seen. Furthermore, it is not known if larger BPVs (i.e. >21 mm) will fracture consistently with valvuloplasty balloons sized 1 mm larger than the labelled BPV size, as the force exerted by larger balloons in larger prostheses may be somewhat different than in small valves, and this has not been systematically tested clinically or on the bench.

Figure 3: A: Baseline Appearance of 23 mm Magna BPV

Article image

Timing of BVF

Whether BVF is optimally performed before or after implantation of the TAVR prosthesis is not known. There are theoretical advantages to both strategies. If BVF is performed first followed by TAVR implantation, this may allow for confirmation that the BPV can be fractured prior to TAVR implantation, which in theory may allow for selection of a larger TAVR prosthesis. However, bench testing and clinical experience have demonstrated consistent success in fracturing most BPVs, thus it does not appear to be necessary to ensure BVF is successful prior to THV implantation.

Some operators have preferred to perform BVF prior to VIV TAVR to avoid the theoretical concern of subclinical damage to the prosthetic leaflets during high-pressure balloon inflation, which might impact long-term THV durability. Performing BVF before VIV TAVR has the potential benefit of sparing the THV the high-pressure balloon inflation. However, it should also be noted that balloon-expandable THV leaflets are subjected to pressure at the time of crimping, and at the time of implantation regardless of whether BVF is performed. Furthermore, as noted above, in self-expanding THV valves, BVF should be performed by placing the balloon shoulder below the level of the supra-annular valve leaflets to avoid any possibility of leaflet injury. At this time there are no robust long-term data regarding THV durability after BVF. The 100 % long-term survival in one case series (n=10) is a promising sign that patients do well after BVF. However, haemodynamic and quality of life data are not available in this cohort.

It is also not known whether there are any differences between procedural outcomes of BVF performed before or after VIV TAVR. However, some data suggest that there may be some disadvantage to performing BVF first, especially when a balloon-expandable THV is used. In bench testing, if a balloon-expandable THV was implanted nominally within a fractured BPV using standard inflation, the compliant delivery balloon was not sufficient to fully expand the TAVR prosthesis, and a notable constraint remained after the TAVR valve was implanted.21 A final, high-pressure balloon inflation was necessary to fully expand the balloon-expandable TAVR prosthesis. Interestingly, when self-expanding TAVR valves were implanted within fractured BPVs, there appeared to be sufficient radial strength to re-expand the fractured BPV and achieve nominal deployment diameter, without a final high-pressure balloon inflation.21

Haemodynamic data may also support the strategy to perform VIV TAVR prior to BVF. In the currently published series (n=30), the final haemodynamics when VIV TAVR was performed first, followed by BVF, appear to be superior to the results when BVF was performed first (Table 2). Although patients treated with BVF first (n=15) had a similar pre-procedural gradient and EOA to those who underwent TAVR first (n=15), the final mean gradient and EOA in the BVF-first group were less favourable compared with the haemodynamics in the TAVR-first group (14 mmHg and 1.4 cm2 vs 7 mmHg and 1.7 cm2, respectively). However, this comparison might be confounded by differences in the valves that were fractured or the THV selected for VIV TAVR. In the BVF-first group, 13 of 15 patients (87 %) were treated with a balloon-expandable THV, with a nominal pressure implantation, and no high-pressure balloon inflation following the THV. These data suggest that a final high-pressure inflation following BVF with a balloon-expandable THV may be necessary to optimise THV expansion and procedural results. It is important to interpret these findings with caution, as the total number of patients in each group is small and the results are subject to confounding.

Longer-term Outcomes

There are some available data regarding haemodynamic durability at 1 month following BVF. In a series of 18 patients who underwent BVF prior to VIV TAVR, the baseline mean transvalvular gradient and EOA were 42.8±17.0 mmHg and 0.8±0.3 cm2, respectively, which improved to 8.1±3.6 mmHg and 1.96±0.58 cm2, respectively, after VIV TAVR and BVF. At 1 month, mean gradient and EOA by echocardiography were 12.7±5.0 mmHg and 1.64±0.3 cm2, respectively, which were not statistically different to the final procedural haemodynamic measurements. Long-term follow-up of these patients is ongoing.

In contrast to concerns that BVF may result in impaired TAVR durability, it is also possible that BVF actually improves durability, considering that a high residual gradient after VIV TAVR is associated with worse outcomes. Poor expansion of prosthetic leaflets has been associated with early BPV failure, as leaflet folding results in stress and strain points on the leaflets, which leads to leaflet calcification and fibrosis, and can accelerate BPV degeneration.28 In theory, by optimising THV expansion and reducing the transvalvular gradient, BVF might decrease leaflet stress and degeneration and improve long-term THV durability. If this is indeed the case, then all patients undergoing VIV TAVR might benefit from BVF, irrespective of the size of their BPV or the residual gradient following THV implantation. Ultimately, comparisons of long-term outcomes between patients who undergo BVF before or after VIV TAVR first will be important to understand the optimal BVF strategy.

Figure 4: A: Baseline Haemodynamics Prior to VIV TAVR

Article image

Table 2: Combined Clinical Cases of BVF and VIV TAVR

Article image

Haemodynamic Support

Due to concern that BVF may result in aortic root injury or coronary artery obstruction, some operators have preferred to perform BVF only in the setting of full haemodynamic backup with ECMO (Table 2). To date, no published reports of aortic root injury or haemodynamic collapse attributable to BVF exist. As clinical experience with BVF has accumulated, it appears that full haemodynamic backup is not routinely necessary. We do not recommend routine use of ECMO during these cases, especially as ECMO requires additional further large-bore vascular access, further exposing the patient to potential vascular complications. However, in certain situations ECMO may be beneficial, such as a patient with very high risk for coronary occlusion with VIV TAVR and BVF.

Future Directions

The initial bench testing and clinical experience with BVF and VIV TAVR establishes a firm foundation for this procedure in patients with failed aortic BPVs. However, there are many unanswered questions relating to this novel technique. Whether BVF has an impact on the survival of patients who are at risk of PPM following VIV TAVR remains to be seen. Further data are needed as to the quality of life benefit that is gained from VIV TAVR with BVF compared with VIV TAVR alone. The feasibility of BVF in patients with larger BPVs (>21 mm labelled BPV size) has not been well studied, and whether this will improve the procedural results or outcomes for these patients is equally unknown. The safety margins for performing BVF in patients at risk of coronary obstruction and aortic root injury are not fully understood and warrant further study. Finally, it remains to be seen whether BVF can be performed safely and successfully, and with benefit, in conjunction with VIV procedures in the mitral, pulmonary or tricuspid positions. As our experience grows, the indications for and technique of BVF will undoubtedly continue to be refined.

References

  1. Adams DH, Popma JJ, Reardon MJ. Transcatheter aortic-valve replacement with a self-expanding prosthesis. N Engl J Med 2014;371:967–8.
    Crossref | PubMed
  2. Leon MB, Smith CR, Mack M, et al. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 2010;363:1597–607.
    Crossref | PubMed
  3. Smith CR, Leon MB, Mack MJ, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 2011;364:2187–98.
    Crossref | PubMed
  4. Leon MB, Smith CR. Transcatheter aortic-valve replacement. N Engl J Med 2016;375:700–1.
    PubMed
  5. Balsam LB, Grossi EA, Greenhouse DG, et al. Reoperative valve surgery in the elderly: predictors of risk and long-term survival. Ann Thorac Surg 2010;90:1195–200.
    Crossref | PubMed
  6. Kaneko T, Loberman D, Gosev I, et al. Reoperative aortic valve replacement in the octogenarians-minimally invasive technique in the era of transcatheter valve replacement. J Thorac Cardiovasc Surg 2014;147:155–62.
    Crossref | PubMed
  7. Dvir D, Webb JG, Bleiziffer S, et al. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA 2014;312:162–70.
    Crossref | PubMed
  8. Webb JG, Mack MJ, White JM, et al. Transcatheter aortic valve implantation within degenerated aortic surgical bioprostheses: PARTNER 2 valve-in-valve registry. J Am Coll Cardiol 2017;69:2253–62.
    Crossref | PubMed
  9. Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation 1978;58:20–4.
    Crossref | PubMed
  10. Blais C, Dumesnil JG, Baillot R, et al. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation 2003;108:983–8.
    Crossref | PubMed
  11. Koene BM, Soliman Hamad MA, Bouma W, et al. Impact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement. J Cardiothorac Surg 2013;8:96.
    Crossref | PubMed
  12. Dayan V, Vignolo G, Soca G, et al. Predictors and outcomes of prosthesis-patient mismatch after aortic valve replacement. JACC Cardiovasc Imaging 2016;9:924–33.
    Crossref | PubMed
  13. Azadani AN, Jaussaud N, Matthews PB, et al. Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses. Interact Cardiovasc Thorac Surg 2010;11:70–7.
    Crossref | PubMed
  14. Simonato M, Azadani AN, Webb J, et al. In vitro evaluation of implantation depth in valve-in-valve using different transcatheter heart valves. EuroIntervention 2016;12:909–17.
    Crossref | PubMed
  15. Simonato M, Webb J, Kornowski R, et al. Transcatheter replacement of failed bioprosthetic valves: large multicenter assessment of the effect of implantation depth on hemodynamics after aortic valve-in-valve. Circ Cardiovasc Interv 2016;9:e003651.
    Crossref | PubMed
  16. Zenses AS, Mitchell J, Evin M, et al. In vitro study of valve-in-valve performance with the CoreValve self-expandable prosthesis implanted in different positions and sizes within the Trifecta surgical heart valve. Comput Methods Biomech Biomed Engin 2015;18(Suppl 1):2086–7.
    Crossref | PubMed
  17. Azadani AN, Reardon M, Simonato M, et al. Effect of transcatheter aortic valve size and position on valve-in-valve hemodynamics: an in vitro study. J Thorac Cardiovasc Surg 2017;153:1303–15.
  18. Nielsen-Kudsk JE, Christiansen EH, Terkelsen CJ, et al. Fracturing the ring of small mitroflow bioprostheses by high-pressure balloon predilatation in transcatheter aortic valve-in-valve implantation. Circ Cardiovasc Interv 2015;8:e002667.
    Crossref | PubMed
  19. Brown SC, Cools B, Gewillig M. Cracking a tricuspid perimount bioprosthesis to optimize a second transcatheter sapien valve-in-valve placement. Catheter Cardiovasc Interv 2016;88:456–9.
    Crossref | PubMed
  20. Tanase D, Grohmann J, Schubert S, et al. Cracking the ring of Edwards Perimount bioprosthesis with ultrahigh pressure balloons prior to transcatheter valve in valve implantation. Int J Cardiol 2014;176:1048–9.
    Crossref | PubMed
  21. Allen KB, Chhatriwalla AK, Cohen DJ, et al. Bioprosthetic valve fracture to facilitate transcatheter valve-in-valve implantation. Ann Thorac Surg 2017;104:1501–8.
    Crossref | PubMed
  22. Chhatriwalla AK, Allen KB, Saxon JT, et al. Bioprosthetic valve fracture improves the hemodynamic results of valve-in-valve transcatheter aortic valve replacement. Circ Cardiovasc Interv 2017;10:e005216.
    Crossref | PubMed
  23. Nielsen-Kudsk JE, Andersen A, et al. High-pressure balloon fracturing of small dysfunctional Mitroflow bioprostheses facilitates transcatheter aortic valve-in-valve implantation. EuroIntervention 2017;13:e1020–5.
    Crossref | PubMed
  24. Johansen P, Engholt H, Tang M, et al. Fracturing mechanics before valve-in-valve therapy of small aortic bioprosthetic heart valves. EuroIntervention 2017;13:e1026–31.
    Crossref | PubMed
  25. Dvir D, Webb J, Brecker S, et al. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry. Circulation 2012;126:2335–44.
    Crossref | PubMed